A novel method to assess the incident angle and the LET of protons using a compact single-layer Timepix detector
Published (publisher metadata verified)
Abstract / research summary
This study presents a method to evaluate the incident angle and linear energy transfer (LET) of protons using a compact single-layer hybrid pixel detector—Timepix. The study explores correlations between the track shape and the proton incidence angle for varying proton energies (89 MeV and 226.7 MeV) and angles (0° to 75°). Monte Carlo simulations using TOPAS were used to validate the experimental results. The detector, equipped with a 300 μm thick silicon sensor, was irradiated with a mono-energetic proton beam and the resulting cluster shapes analyzed to extract information about incident angles and LET values. A statistical model based on cluster elongation and deposited energy allowed for effective classification and angle estimation. The findings demonstrate the feasibility of using a single-layer Timepix device for angular and LET characterization in proton fields, which could support particle therapy quality assurance and beam monitoring applications.
Text supplied in this site's publication record. Consult the full paper for methods, experimental conditions and limitations.
Manuscript version and redistribution rights have not been independently confirmed for all files. An accessible download is not proof of an open-access licence.
Cite this publication
Bibliographic reference and permanent page on this site
Permanent page: https://cristinaoancea.com/publications/90
Bibliographic metadata verified against the Crossref registration record. Abstract text and manuscript version are not verified by this check.