Configuration of Timepix3 read-out parameters for spectral measurements in proton therapy applications

Paulina Stasica-Dudek, Cristina Oancea, Carlos Granja, Konrad Guguła, Jan Jakubek, Renata Kopeć, Dawid Krzempek, Michal Matous, Antoni Rucinski, Marzena Rydygier, Jan Gajewski
Physica Medica
2025

Published (publisher metadata verified)

DOI: 10.1016/j.ejmp.2024.104885

Abstract / research summary

Purpose: With the increasing use of proton therapy, there is a growing emphasis on including radiation quality, often quantified by linear energy transfer, as a treatment plan optimization factor. The Timepix detectors offer energy-sensitive particle tracking useful for the characterization of proton linear energy transfer. To improve the detector's performance in mixed radiation fields produced in proton therapy, we customized the detector settings and performed the per-pixel energy calibration. Methods: The detection threshold and per-pixel signal shaping time (I krum ​ current) were customized, and energy calibration was performed for MiniPIX Timepix3. The detector calibration was verified using α source and clinical proton beams, as well as Monte Carlo simulations. The effects on the detector's performance, in terms of spectral saturation and pixel occupancy, were evaluated. Results: Measurements with proton beams showed a good agreement with simulations. With the customized settings, the measurable energy range in the detector data-driven mode was extended, and the signal duration time was reduced by 80%, while the yield of pixel time occupancy reduction depends on the number of occupied pixels. For performed measurements with proton beams, the number of occupied pixels was further reduced up to 40% due to the increased threshold. Conclusions: Customized detector configuration of the Timepix3 detector allowed for reduced pixel occupancy and mitigation of signal saturation in a data-driven mode without significantly interfering with the energy deposition measurement. The presented approach enables the extension of the operational range, including higher intensities and mixed-radiation fields in particle radiotherapy environments.

Text supplied in this site's publication record. Consult the full paper for methods, experimental conditions and limitations.

Manuscript version and redistribution rights have not been independently confirmed for all files. An accessible download is not proof of an open-access licence.

Cite this publication

Bibliographic reference and permanent page on this site

Paulina Stasica-Dudek, Cristina Oancea, Carlos Granja, Konrad Guguła, Jan Jakubek, Renata Kopeć, Dawid Krzempek, Michal Matous, Antoni Rucinski, Marzena Rydygier, Jan Gajewski (2025). Configuration of Timepix3 read-out parameters for spectral measurements in proton therapy applications. Physica Medica. https://doi.org/10.1016/j.ejmp.2024.104885

Permanent page: https://cristinaoancea.com/publications/107

Bibliographic metadata verified against the Crossref registration record. Abstract text and manuscript version are not verified by this check.